Through our global network of testing experts and analytical equipment including chromatography (HPLC, GC, GC/MS) and atomic
absorption spectroscopy (AAS, GFA, FIAS), Our goal is to provide test services as efficiently
as possible to maximize our customers' profits. For more information about our services,
contact one of our experts today.
Note: this service is for Research Use Only and Not intended for clinical use.
Using differential scanning calorimeter (DSC), we can study the phase transition of inorganic materials, the melting and crystallization process of polymer materials, the polycrystalline phenomenon of drugs and so on. DSC 204 F1 Phoenix is used to study the thermal effects of materials at controlled temperature. It provides liquid nitrogen, mechanical refrigeration, air compression and cooling cup for different cooling methods. The Tau-sensor has good separation ability to the overlapping thermal effects, the µ-sensor has a high level of calorimetric sensitivity, and can be selected according to different detection needs.
Applications
Phase transition of inorganic materials.
Melting and crystallization of polymer materials.
Glass transition.
Oxidative stability.
Polycrystalline phenomenon of drugs.
Solid/liquid ratio of foods such as fats and oils.
Melting point of metals and alloys.
Widely used in plastics, rubber, coatings, adhesives, pharmaceuticals, petrochemical and other fields.
Specifications
Specifications-DSC 204 F1 Phoenix
Temperature Range
-180°C to 700°C
Heating Rates
0.001 K/min to 200 K/min
Cooling Rates
max. 200 K/min
Compressed Air Cooling
RT to 700°C
Intracooler for the Extended Range
-85°C to 600°C
Automatically Controlled LN2 Cooling
-180°C to 700°C
Indium Response Ratio
>160 mW/K
Peak Height [mW]/Peak with at 50% Height [K] of Indium melting peak
Automatic Sample Changer
for 192 samples and reference crucibles (optional)
Features
Determination of onset, peak, inflection and end temperatures
Automatic peak search
Evaluation of crystallization
Comprehensive glass transition analysis
Automatic baseline correction
Specific heat determination
Sector-specific software solutions for conformance to e.g. GLP and GMP